19252 Wrist strip /Foot-grounding test device

Calibration procedure:

The resistance test by W/S tester must be over 740K次(working resistance as 750K次-10.5M次), however, highest resistance must be less than 11M次, and resistance tested in shoe cover must be over 740K次( working resistance is 750K次-35M次), however it must be less than 40M次, and resistance tolerance is 20%.

Wrist test margin: 750K次-10.5M次
Shoe cover test margin: 750K次-35M次
Resistance tolerance (㊣20%) Test output Resistance tolerance (㊣20%) Test output
≒740K次 Red (FAIL LO) ≒740K次 Red (FAIL LO))
750K次 Green (PASS (Normal)) 750K次 Green (PASS (Normal))
10. 5M次 Green (PASS (Normal)) 35M次 Green (PASS (Normal))
≡11.5M次 Red (FAIL HI (Fail Hi)) ≡40M次 Red (FAIL HI (Fail Hi))

It shall be recommended to recalibrate the measurement instrument once a year from calibration time last time.

Instruction:

MAXCLEAN M-FWS-1 tester can be touchable and used as 3 status, which is ideal for fast and multiple test for ESD grounding device. This tester shall be used as one of calibration instruments according to the calibration standard inclusive of daily inspection of technical requirement and program described as 6.1.3.2 in ANSI ESD S20.20. Combo tester integrated with design for unique dual-testing circuit, which not only improve test speed, but also does not need any single wrist strip/foot panel tester. Furthermore, FWS-1 configures with 750K次-10M次 circuit used to test wrist strip and 750K次-35M次 circuit used to test shoe cover, FWS-1tester is ideal for testing of wrist strip and shoe cover. Test parameter has been preset in factory, however it can be used only after one needed by you coordinates with test parameter. Here FWS-1 tester is easily operable, green indicator stands for normal operation status, red one and sound for higher or lower resistance

Test for wrist strip

Test can be used to calibrate if there is switched on among operator, wrist strip and grounding lead, see diagram I.

A. Plug end shall be put into frontal socket of tester when wrist strip is worn.
B. Switch shall be turned on WRIST CORD.
C. Tester shall be operated after 2-3s if its button is pushed on. Notes: metal part must not be touched by mistake when testing.
D. If green indicator "Normal PASS"LED turns on, it shall indicate normal operation between wrist strip and grounding lead.
E, If "FAIL LO"or"FAIL HI"LED turns on and alarms, inspection shall be done.


Foot-grounding test device

 

 

 

 

 

 

 